Relationship between Patient-Dependent Parameters and Radiation Dose Rates Measured around Patients Undergoing PET/CT Imaging Using 18F-FDG
Alenezi, kahled Soliman, saad Alqahtani, Ahmed . 2018
Objectives first to measure the patient attenuation factor by performing direct dose rate measurements from patients undergoing PET/CT imaging studies using 18F-FDG. The second objective was to study the possible correlation between the measured dose rate constant per unit activity from the patients and their body size metrics; five metrics were tested in this work. The last objective was to measure the patients’ voiding factor. Methods: We have measured dose rates at one meter from 57 patients and noted the patient’ ± 14 proposed by the AAPM, TG-108 report is adequate for radiation protection purposes. There was no statistically significant correlation between the dose rate constant per unit activity and the patient body size metrics. We have measured a patient voiding factor of 0.89 ± 0.06 in comparison with 0.85 recommended by the AAPM. Conclusions: The presented data can be used by medical physicist working in nuclear medicine in formulating more accurate risk estimations resulting from radiation exposure from patients undergoing 18F-FDG PET/CT imaging