A comparison of CsI:Tl and GOS in a scintillator-CCD detector for nuclear medicine imaging
Journal Article
Bugby, Sarah L . 2016
نوع عمل المنشور:
Full paper
رابط المنشور على الويب:
المجلة \ الصحيفة:
Journal of Instrumentation JINST
رقم العدد:
9
رقم الإصدار السنوي:
11
الصفحات:
P09009-P09009
مستخلص المنشور:
A number of portable gamma cameras for medical imaging use scintillator-CCD based detectors. This paper compares the performance of a scintillator-CCD based portable gamma camera with either a columnar CsI:Tl or a pixelated GOS scintillator installed. The CsI:Tl scintillator has a sensitivity of 40% at 140.5 keV compared to 54% with the GOS scintillator. The intrinsic spatial resolution of the pixelated GOS detector was 1.09 mm, over 4 times poorer than for CsI:Tl. Count rate capability was also found to be significantly lower when the GOS scintillator was used. The uniformity was comparable for both scintillators.