Evaluation of the Small-animal Nano Scan PET/CT System using 89Zr

Sassi, 1. Khaled Alzimami, Sitah Alanazi, Magdi Gannam, Ahmad Alanazi, Ibrahim Aljamaz, Suliman Alyanbawi, Basem Alotaibi, Yousif Almalki, Abdelmoneim Sulieman and Salem . 2020

Introduction: The purpose of the present work was to evaluate the imaging characteristics of 89Zr-PET in comparison with those obtained using fluorine-18 Fluorodeoxyglucose (18F-FDG) PET (a gold standard tracer in PET imaging) using a small-animal NanoScan PET/CT scanner.

Methods: The system’s spatial resolution, sensitivity, uniformity, and image quality were measured on a Nano Scan small-animal PET/CT scanner according to the NEMA NU4-2008 protocols. For reconstruction images, we used 2D and 3D reconstruction algorithms. The reconstruction methods included filter back projection (FBP), the ordered subsets expectation maximization (OSEM) algorithm, and the 3D Tera-Tomo algorithm, which are developed for the NanoScan small-animal PET/CT scanner.

Results: The results obtained showed a significant difference in the spatial resolution for 89Zr as compared to 22Na and 18F when using a 2D reconstruction algorithm. Where the spatial resolution values were much enhanced by using the 3D Tera-Tomo reconstruction for each isotope, the Full width at half maximum (FWHM) values were less than 1 for all isotopes at the center of the field of view (FOV). This difference in spatial resolution is dependent on the positron range, energy and the reconstruction method.

Conclusion: The long half-life of 89Zr makes it an ideal positron emitter for performing immuno-PET, which is matched with the biological half-life of intact mAbs. 89Zr can also give several advantages over other long half-life positron emitters in relation to the overall imaging performance because of its relatively short positron range and simpler decay scheme. The values of 89Zr sensitivity that were obtained in the present study were less than those of previous studies

نوع عمل المنشور
Research Article
رقم الانشاء
16
مجلة/صحيفة
Current Medical Imaging
الصفحات
E-pub Ahead of Print
مزيد من المنشورات
publications

Adequate understanding of radiopharmaceutical distribution in the body of the patient has both spatial and temporal characteristics and they are the key factor to consider when planning successful…

بواسطة 2. Khaled Soliman, Ahmed Alenezi, Abdullah Alrushoud, Salman Altimyat, Mousa Bakkari and Hanaa Alshikh
2021
تم النشر فى:
intechopen
publications

ABSTRACT

بواسطة 1. Bakkari M, Alsubaie B, Alenezi A, Soliman K.
2021
publications

Introduction: The purpose of the present work was to evaluate the imaging characteristics of 89Zr-PET in comparison with those obtained using fluorine-18 Fluorodeoxyglucose (18F-FDG) PET (a gold…

بواسطة 1. Khaled Alzimami, Sitah Alanazi, Magdi Gannam, Ahmad Alanazi, Ibrahim Aljamaz, Suliman Alyanbawi, Basem Alotaibi, Yousif Almalki, Abdelmoneim Sulieman and Salem Sassi
2020